1. Introduction
The ability of a cartridge filter to be sterilised in situ within its housing represents a significant operational advantage for many filter users. Steam in place (SIP) avoids the need to use chemical sanitising agents or to compromise the integrity of the filtration installation, minimises operator hands on time and plant downtime and can be easily incorporated into automated production facilities. To achieve reproducible sterilisation conditions (typically a minimum of 121°C for 30 minutes) and to avoid damage to the installed filter cartridge(s), it is important to carefully design and monitor SIP procedures. Factors including steam temperature and quality, condensate removal, differential pressure and cooling cycles need to be considered. The purpose of this document is to detail the recommended methods of SIP for domnick hunter cartridge filters, to avoid accidental damage to cartridges and maximise the steam life of installed filters.
2. General Principles
2.1. Steam Quality
Condensate:
Optimum steam sterilisation of filter cartridges can be achieved using dry steam. Wet steam (steam containing high level of condensed water) will not flow easily through the filter. This resistance to flow will generate increased differential pressure across the filter at elevated temperatures, conditions which cause the maximum stress to the filter cartridge. Condensate should therefore be removed by a manual valve or an automated steam trap, located as close to the filter as possible, to avoid condensate contact with the filter.
Particulate:
Many steam lines suffer from corrosion over a prolonged time period. This can introduce particulate into the steam which will be retained by the filter being steamed. This will accelerate the rate of filter blockage or worse, cause damage to the filter. Pipe corrosion can result in metal fragments which when carried to the filter could puncture the support materials and membrane. To control particulate in steam lines, so protecting the filter cartridge to be steamed, domnick hunter supply a range of in line steam filters. Details of the domnick hunter stainless steel steam filter range are available from your local domnick hunter office.
Chemical additives:
Chemicals are often added to the feed water in steam generators. These chemicals will form part of the steam and so will contact the filter cartridge being steam sterilised. Most standard steam generator additives do not pose any problem for the steam filter or filter cartridge being sterilised. If you have any doubt regarding steam generator additives please contact Technical Support Group, Process Filtration, domnick hunter.
2.2. The Temperature Pressure Relationship
There is a direct relationship between steam temperature and pressure. Increasing the steam pressure facilitates an increase in the temperature of the pressurised steam. The following graph illustrates this relationship.

1 Thermodynamic and Transport Properties of Fluids. SI units by G.F.C Rogers and Y.R Mayhew. Oxford Basil Blackwell 1981 (0631 128913).
As saturated steam is pressurised its temperature increases. To assure sterility, it is typically recommended that a temperature of 121°C be maintained over at least 30 minutes. A pressure of 1.1 barg is required to achieve this temperature (actually equates to 121.8°C). However, 1.0 barg equates to 120.2°C and so would not be sufficient to give a high degree of assurance that sterility has been achieved. It is therefore important that appropriate devices are installed to directly record either steam temperature or the pressure.
2.3. Differential Pressure
Filters should generally be supplied with steam from the upstream side (normal flow direction). Steam flow from the downstream side should be avoided, but may be appropriate in some circumstances for air filters. Reverse flow should not be used for liquid filters. Steam should not be applied to both sides of the filter at the same time as this can potentially trap a pocket of dry air between the steam supplies, reducing the heat transfer to the filter and therefore negatively impacting the steam sterilisation conditions. During normal steam sterilisation cycles, as steam is applied to the filter housing, air is purged from the system such that saturated steam contacts and heats all the filter and housing surfaces. As the steam heats and subsequently flows through the cartridge filter, there will be an inherent resistance to that flow. The amount of resistance to the steam flow is measured as a drop in pressure across the filter and is called the differential pressure.

Steam pressure can be expressed in two forms:
1. Absolute pressure, which includes atmospheric pressure.
Therefore sterilisation would require 2.1 bar (absolute).
2. Gauge pressure is the pressure above atmospheric pressure.
Therefore sterilisation would require 1.1 bar (gauge).
The differential pressure can be influenced by many factors including the micron rating of the cartridge filter, the degree of blockage due to retained contamination, the influence of condensate and the rate of steam flow through the cartridge filter. Whilst filter cartridges can withstand high differential pressures (up to 5 barg) at ambient temperature, it is important to limit differential pressure at elevated temperature. When a cartridge filter is heated to 121°C or higher, this procedure places significant stress on the unit. During the SIP process the physical properties of the cartridge filter components are weakened at elevated temperature (e.g. increased plasticity of polymers). It is therefore important that at temperatures of 121°C or higher, the differential pressure across the filter cartridge should not exceed 0.3 barg. Differential pressure above this level may lead to filter damage. Consideration of process parameters to limit differential pressure during SIP is therefore critical and includes:
- Ensuring efficient condensate removal during the entire SIP cycle
- Minimising steam flow rates through the cartridge filter.
- Minimising the downstream pipework steamed through the filter cartridge
- Control and measurement of the temperature and differential pressure during filter cartridge SIP.
- Removal, post SIP, of as much of the steam as is practical to reduce condensate formation the cartridge filter. This is particularly relevant for hydrophobic gas filter cartridges.
- Controlled cooling of filter cartridges post SIP, to ensure hydrophobic filter cartridges are not blinded by condensate.
- Controlled cooling of filter cartridges post SIP to prevent potential thermal shock by crash cooling or cold product introduction.
2.4. Filter Housing Installation
The following considerations should be given to the installation of filter housings that will be steam sterilised: - The installation should ensure that condensate cannot accumulate in the housing. Housings or upstream pipework should therefore be fitted with condensate drains. Condensate removal is particularly important when steam sterilising hydrophobic gas filter cartridges. domnick hunter manufacture a range of gas filter housings which incorporate plenum chamber. This upstream chamber collects steam condensate and inhibits contact with the hydrophobic filter cartridge. The plenum chamber is one of the specific design features that differentiate domnick hunter gas filter housings from liquid filter housings. - The cartridge and housing should be in a normal orientation (vertical, with the open end of the cartridge pointing downwards). - The pipework downstream of the filter housing should be as short as possible. domnick hunter recommend a maximum of 100cm. Tanks downstream of the filter should always be steam sterilised separately. - Pressure gauges capable of accurate reading over the minimum range of 0-3bar at the SIP temperature should be installed upstream and down stream of the filter housing, to allow differential pressure measurement. - The pipework should be positioned at a slight incline allowing gravity to collect the steam condensate at a steam drain or steam trap. - Length of pipeline, relative to steam flow rate, may result in pressure loss. The length of steam supply pipework should be kept to a minimum.
2.5. Condensate Removal
Whilst saturated steam behaves as a gas and so flows easily through filters, contact with any cool surface (e.g. stainless steel housings or pipework) will lead to the generation of condensate as the steam cools. In some systems, very large volumes of condensate may be generated due to this process (e.g. multiple cartridge housings, extended pipework run etc). In any case, the removal of condensate from any SIP system is important for a number of reasons: 1. Condensate can ‘blind’ both hydrophilic and hydrophobic membranes to steam flow, potentially leading to filter damage due to the development of high differential pressures across the membrane at high temperature. 2. Condensate will be at a temperature below the required steam sterilisation temperature. It is therefore important to remove condensate to ensure effective steam sterilisation. 3. Hydraulic shock may occur due to ‘slugs’ of steam being forced through the line. In designing process systems, the effects of condensate generation can also be minimised by ensuring that housings are not located at the bottom of long pipe runs, or that these are steam sterilised separately from the filter.
2.6. Steam Traps
As discussed above, the removal of condensate from systems is important to ensure efficient steam sterilisation. A thermodynamic steam trap or condensate trap is a mechanical valve to remove condensate. They are located at points upstream and downstream of the filter where condensate would collect (i.e. low points in the system). These steam traps work on the principle that as condensate collects the temperature at the trap falls below that required for effective steam sterilisation (121°C). At this point the valve opens, drains the condensate and draws in live steam from the steam supply. Steam traps can be replaced by a manual valve that is left slightly open, but requires control by a skilled operator to ensure effective condensate removal without draining excessive steam.
2.7. Integrity Testing
Steam sterilisation under carefully controlled conditions is an accepted process operation for filter cartridges. However, this does represent an aggressive process condition and can lead to filter damage should steam sterilisation conditions deviate from controlled norms. It is therefore recommended that filter cartridges are integrity-tested in-situ after steam sterilisation and before use. For advice on in-situ integrity testing of filter cartridges please refer to domnick hunter.
2.8. Cartridge Filter Steam Life
domnick hunter cartridge filters undergo extensive steam sterilisation verification trials during their development. During these trials a number of filter cartridges from several production lots are steam sterilised at 121-142°C depending on the filter type. After each 30-minute steam sterilisation cycle the filter under test is integrity tested. Testing continues until the cartridge fails integrity test and the steam life for that cartridge type is then defined from this data, incorporating a safety margin. Recommended maximum steam life for domnick hunter filter cartridges at a range of temperatures is shown in Table 1. It should be noted that these are only provided as guidelines as actual filter life will depend on individual steam sterilisation conditions. 
3. Practical SIP
The following sections in this guide provide step by step procedures for developing SIP protocols in three operational conditions: 1. Liquid filters steamed in the forward direction.
The steam flows in the same direction as the liquid product (from outside to inside of the filter cartridge). The filter has a media or membrane, which is hydrophilic. 2. Gas filters steamed in the forward direction.
The steam flows in the same direction as the gas product (from outside to inside of the filter cartridge). The filter has a media or membrane which is hydrophobic. 3. Gas filters steamed in the reverse direction.
The steam flows in the opposite direction (inside to outside of the filter cartridge) to the designed direction of flow for the gas product. The filter has a media or membrane which is hydrophobic. Each procedure has a system diagram and valve sequence table to illustrate the step by step procedure. The procedures represent ideal systems for SIP, which may not be identical to existing systems. For recommendations, modifications or more information regarding these procedures please contact the domnick hunter Technical Support Group.
4. Steam-in-Place Procedures
4.1 Forward Steam-in-Place Procedure Liquid Filter Applications
1. Set the valves to positions indicated for pre Steam-In-Place.
2. Drain the product from the filter system and associated pipework. Opening valve V5 will aid this process.
3. Open valve V1 and allow the steam condensate to drain until the steam trap below valve V3 closes. Close valve V9.
4. Slowly open V3 allowing steam into the system: this will flow across the filters and through valve V4 & V5. This will allow the heating of the housing, the filters and associated pipework without generating a significant differential pressure across the filters. Note: The steam trap below valve V3 & V4 will receive the condensate and will repeatedly open and close.
5. When ‘live’ steam flows from valve V5, close valve V5. This will direct the steam through the heated filter. close valve V10.
6. Observe the pressure gauges P1 and P2 and control the steam flow rate at valve V3 to ensure the differential pressure does not exceed 0.2 - 0.3 barg.
7. When the steam trap below valve V6 closes, the steam pressure will begin to rise.
8. Ensure the steam pressure/temperature does not exceed the maximum allowable pressure/temperature for the cartridge type being steamed. If reading from pressure gauges it is recommended the maximum steam pressure is 2.0 barg in the forward direction.
9. Steam sterilise the cartridges for 30 minutes ensuring the conditions stated in steps 5 to 7 are followed. The valves should now be in positions indicated for Steam-In-Place.
10. On completion of the Steam-In-Place cycle, close V4, V6, V3 and V1 in that order.
11. Slowly open V10 to release the steam pressure from the filter system and associated pipework. When the pressure on P2 reads 0.1 barg pressure close valve V10. Fully open valve V9 to release the remaining steam pressure from the filter system. When the pressure on P1 reads 0.1 barg pressure, close valve V9.
12. Allow the system to cool for 30 minutes. The valves should now be in the positions indicated for post Steam-In-Place.
NOTES:
1. A double downstream valve (V7, V8) is recommended so that under the cartridge steaming protocol the valves sealing faces of V7 can be effectively sterilised. The sealing valve faces of V8 can be similarly sterilised when the tank is steamed. When steam sterilising the tank, V7 would be closed and V6 and V8 open. Normally the tank would be steamed separately before steaming the filter. If the filter is steamed before steaming the tank it is recommended that valve V7 is closed in the post Steam-In-Place settings to maintain sterility. The valve V7 must be closed during Step 10.
2. Valve V7 should be installed horizontally and valve V6 / steam trap installed immediately downstream of V7.
3. All drains should be fitted vertically to allow liquid removal.
4. Large volume downstream systems should not be steam through the filter; e.g. when steaming process tanks a secondary steam supply should be used.
4.2 Forward Steam-in-Place Procedure Air Filter Applications
1. Set the valves to positions indicated for pre Steam-In-Place.
2. Open valve V1 and allow the steam condensate to drain until the steam trap below valve V3 closes.
3. Slowly open V3 allowing steam into the system: this will flow across the filters and through valve V4 & V5. This will allow the heating of the housing, the filters and associated pipework without generating a significant differential pressure across the filters. Note: The steam trap below valve V3 & V4 will receive the condensate and will repeatedly open and close.
4. When ‘live’ steam flows from valve V5, close valve V5. This will direct the steam through the heated filter.
5. Observe the pressure gauges P1 and P2 and control the steam flow rate at valve V3 to ensure the differential pressure does not exceed 0.2 - 0.3 barg.
6. When the steam trap below valve V6 closes, the steam pressure will begin to rise.
7. Ensure the steam pressure/temperature does not exceed the maximum allowable pressure/temperature for the cartridge type being steamed. If reading from pressure gauges it is recommended the maximum steam pressure is 3.0 barg in the forward direction.
8. Steam sterilise the cartridges for 30 minutes ensuring the conditions stated in steps 5 to 7 are followed. The valves should now be in positions indicated for Steam-In-Place.
9. On completion of the Steam-In-Place cycle, close V4, V6, V3 and V1 in that order.
10. Fully open V5 to flash-dry the filter (or 10a). 10a. Open V2 to allow compressed air into the system. The pressure of the air should be no more than 0.5 barg above the steam pressure.
11. Allow the system to cool for 15 minutes, then close V5 (flash-dry only). The valves should now be in the positions indicated for post Steam-In-Place.
NOTES:
1. A double downstream valve (V7, V8) is recommended so that under the cartridge steaming protocol the valves sealing faces of V7 can be effectively sterilised. The sealing valve faces of V8 can be similarly sterilised when the tank is steamed. When steam sterilising the tank, V7 would be closed and V6 and V8 open. Normally the tank would be steamed separately before steaming the filter. If the filter is steam before steaming the tank it is recommended that valve V7 is closed in the post Steam-In-Place prior to maintain sterility (step 9).
2. Valve V7 should be installed horizontally and valve V6 / steam trap installed immediately downstream of V7.
3. All drains should be fitted vertically to allow liquid removal.
4. Large volume downstream systems should not be steam through the filter; e.g. when steaming process tanks a secondary steam supply should be used.
5. Installation/use of domnick hunter air filter housings which utilise a plenum chamber are recommended, as this facilitates the collection and drainage of condensate.
6. When steaming in the reverse direction please contact domnick hunter for guidance.
4.3 Reverse Steam-in-Place Procedure Air Filter Applications
1. Set the valves to positions indicated for pre Steam-In-Place.
2. Open valve V1 and allow the steam condensate to drain until the steam trap below valve V2 closes.
3. Slowly open V2 allowing steam into the system.
4. Observe the pressure gauges P1 and P2 and control the steam flow rate at valve V2 to ensure the P across the filter does not exceed 0.1 barg. If the P exceeds 100mbar stop the sterilisation procedure and rectify the cause of the pressure drop before proceeding with the sterilisation routine.
5. When line steam flows from valve V6, close valve V6. When the steam trap below valve V5 closes, the steam pressure will begin to rise.
6. Ensure steam pressure/temperature does not exceed the maximum allowable pressure/temperature for the cartridge type being steamed. Continue to monitor the P using gauges P1 and P2. If the differential pressure exceeds 100mbar stop the sterilisation procedure and rectify the cause of the pressure drop before proceeding with the sterilisation routine.
7. Steam sterilise the cartridges for 30 minutes. The valves should now be in the positions indicated for Steam-In-Place. (This method is typically suited to single cartridge systems).
8. On completion of the steam cycle time, close V5, V2, V1 in that order and then rapidly open V6 to flash dry the filter (or 8a). 8a. Isolate steam supply and open V7 to allow air into the system. The pressure of the air should be no more than 0.5 barg above the steam pressure.
9. Allow the system to cool for 15 minutes then close V6, the valves should now be in positions indicated for post Steam-In-Place.
NOTES:
1. A double downstream valve (V4, V3) is recommended so that under the cartridge steaming protocol the valve sealing faces of V4 can be effectively sterilised - the sealing valve faces of V3 can be similarly sterilised when the tank is steamed. When steam sterilising the tank V4 should be closed and valve V3 open. To ensure adequate drainage at V3, there should be a slight fall in the pipework between V3 and the tank. As the same steam supply is used to sterilise the tank as well as the filter. A steam filter of a suitable flow rate capacity must be selected.
2. All drains should be fitted vertically to allow liquid removal.
3. There should be a ‘fall’ in the pipework from V3 so that liquid can drain into the tank.
4. Large volume downstream systems should not be steamed through the filter; e.g. when steaming process tanks a secondary steam supply should be used.
5. Installation/use of domnick hunter air filter housings which utilise a plenum chamber are recommended, as this facilitates the collection and drainage of condensate.
6. Valve V7 should be placed as close as possible to the filter housing. If long lengths of pipe are present, the initial steam flows will be high as the downstream volume is pressurised. This high steam flow can change the cartridge.
5. Technical Support Group activities
domnick hunter have a trained team of scientists and engineers available to answer questions regarding the technical capabilities of our products, to assist in the selection and design of appropriate filtration systems and to provide user training programs. The following services can be delivered both on site and in-house;
- filterability testing to optimise filter system design
- advice on the development of integrity testing, steam sterilisation and clean in place procedures
- development of validation procedures
- troubleshooting
- facility audits to ensure continued optimisation of filter use
- operator training including filtration theory, filter system design and management, validation, etc.
For more information on any of the above support services please contact your local domnick hunter representative. email: tsg@domnickhunter.com